Operation, Bionix envirobath chamber, Heating circulator bath operation – MTS Bionix EnviroBath User Manual

Page 35: Process fluids, Set limits before testing, Elevated temperature operation

Advertising
background image

Bionix® EnviroBath™ Chamber

Operation

35

Operation

Bionix EnviroBath

chamber

Other than the hose connections and closing and latching the door before filling
the chamber with fluid, all operation is performed using the heating circulator
bath.

Heating circulator

bath operation

A manual describing operation of the heating circulator bath is provided by the
vendor.

Process fluids

Use only recommended fluids:

Water

Saline

Protein serum

Set limits before

testing

Limits should be set such that grips and fixtures cannot damage the EnviroBath
chamber during setup and testing.

For electromechanical test frames, limits are usually set by positioning physical
switches to stop crosshead travel when the limit is reached. For more
information, see the individual frame product manual.

For servohydraulic test frames, limits are usually set in the control software. For
more information, see setting detectors in the controller software manual for
additional information.

Elevated temperature

operation

For elevated temperature operation information, see circulating bath vendor
documentation

Fluids used in the EnviroBath chamber can reach boiling point.

Boiling fluid can cause personal injury and equipment damage.

When operating the EnviroBath chamber at elevated temperature, it is the
customer’s responsibility to ensure the temperature set point is below the boiling
point of the process fluid.