Appendix, Additional digital i/o information, Appendix 69 – MTS Insight Material Testing Systems User Manual

Page 69: Additional digital i/o information 69

Advertising
background image

Additional Digital I/O Information

MTS Insight™ Material Testing System

Appendix

69

Appendix

Additional Digital I/O Information

The digital inputs have an MOCD223 optical isolator with 2.7 Kohm, ½ Watt
series resistor. To reliable turn on they need 1 mA of current. This means the
minimum input high voltage is 4.0 V DC. The maximum input voltage is 28.0 V
DC. The device should be off for input voltages less than 1.0 V.

The digital outputs are implemented by an AQV252G PhotoMOS relay with a
0.75 Amp poly fuse in series. Although the device is rated at 60 V peak but it is
recommended that a maximum of 48 V be applied. If the load is highly inductive,
such as a relay coil, an appropriate snubber network should be used near the coil
terminals to prevent large flyback voltages from exceeding the device ratings.

Insight Controller

User
Digital
Output

User
Digital
Input

From Logic
Controller

To Logic
Controller

Optical
Isolator

PhotoMOS
Relay

Advertising