Grounding the console and chassis, Grounding the console and chassis 62 – MTS Model FlexTest SE User Manual

Page 62

Advertising
background image

Models FlexTest® IIm/GT/SE Controller Hardware

How To Cable the Analog Chassis to the Hydraulic

Installation

62

How To Cable the Analog Chassis to the Hydraulic Control Panel

Connect J50 DC Pwr on the 497.01 Analog Chassis rear panel to J50 on the
497.05 Hydraulic Control Panel rear panel.

Grounding the Console and Chassis

The system will not function correctly if it is not grounded properly.

The power ground and chassis ground terminals are wired together when the
chassis is manufactured.

Remove the jumper between the two grounds and connect each terminal as
follows:

Each chassis ground should be connected to the console with a short ground
cable (PN 037-766-102).

Each power ground should be connected at a common location on the power
panel (PN 037-766-1xx).

The power panel ground should be connected to the console (PN 037-766-
101).

The system ground cable connects the console to other system components (PN
005-402-3xx).

J50

J1

497.05 Hydraulic Control Panel

497.01 Analog Chassis

PN 046-239-1xx

Advertising