Model 493.21b (duc b) jumpers, Model 493.21b (duc b) jumpers 92, Force/strain sensor connections – MTS Model FlexTest SE User Manual

Page 92: Active guard

Advertising
background image

Models FlexTest® IIm/GT/SE Controller Hardware

Model 493.21B (DUC B) Jumpers

FlexTest GT Controller Connections

92

Force/strain sensor

connections

Force and strain sensors require a DC conditioning mezzanine card.

Model 493.21B (DUC B) Jumpers

The following jumpers are for the Model 493.21B Digital Universal
Conditioner). The jumpers are labeled W instead of X; the W jumpers are solder
pads. The following jumper descriptions are from the 493.XX Module Setup
configuration drawing (PN 049-336-301).

Active guard

Jumper W1 enables or disables an active guard for the sensor cable.

Active guard is usually not required. In particular environments, active guard
drive may reduce transducer noise.

4

12

2

15

1

10

7

3

5

13

9
8

11
14

6

E4

E2

E1

E7

E3

E5

Inner

Guard

4

12

2

15

1

10

3

5

13

9
8

11
14

6

Remote Cal

Clk

Data to ID

Data from ID

Sig Gnd

Transducer

ID Module

ID Circuit

Remote Cal

-FB

-FBR

+FB

+FBR

-EX

-EXS

+EX

+EXS

DC Sensor

Backshell

GND

Future

Future

Future

Future

W1

W1

Enable

Disable

Default

Advertising
This manual is related to the following products: