Bunn Electric Brewing System LCR-2 User Manual

Page 18

Advertising
background image

18

36026 120105

MENU SCREEN

ACTION

DESCRIPTION

CAL RIGHT PUMP ?

PULL DISPENSER

CAL RT PUMP VOL

(-) 43mL (+)

CAL RT WTR FLOW

PULL DISPENSER

CAL RT WTR VOL

(-) 36.0OZ (+)

CAL RIGHT SIDE ?

EXIT YES

CAL HOT WTR FLO ?

PULL DISPENSE

CAL HOT WTR VOL

(-) 34.0OZ (+)

XXX CAL -> XXX

(-) TankTemp (+)

Place a 50 mL (#34843.1000)

graduated cylinder under

the Right dispense tip, then

activate the Right Dispenser

Use the (+) or (-) buttons to

enter volume collected

Place a 64 Oz. (2000 mL)

container under the Right

dispense tip, then activate the

Right Dispenser

Use the (+) or (-) buttons to

enter volume collected

NONE

Place a 64 Oz. (2000 mL)

container under the Hot Water

dispense tip, then activate the

Hot Water Dispenser

Use the (+) or (-) buttons to

enter volume collected

Use the (+) or (-) buttons to

adjust temperature reading

The Right Pump will dispense concentrate for

20 seconds, then shut off automatically. Collect

and measure the concentrate dispensed.

Enter the volume of concentrate measured from

the Right Pump in mL(not Oz.) regardless of

Preferred Units selected.

The Right Dispenser will dispense hot water for

20 seconds, then shut off automatically. Collect

and measure the water dispensed.

Enter the volume of the water measured from the

Right dispenser in Oz. (mL)

Scroll down to bypass this screen.

The Hot Water Dispenser will dispense for 20

seconds, then shut off automatically. Collect and

measure the water dispensed.

Enter the volume of the water measured from the

Hot Water Dispenser in Oz. (mL)

Adjust the display reading to match a calibrated

probe inserted next to the Tank Sensor, °F (°C).

If NO - skip to next menu.

Used to set the maximum adjustment the Flavor

Guard feature is allowed to make, [3 - 15%].

Displays the Left Dispenser target RPM and

Conductance Reading, (Diagnostics Only)

Displays the Right Dispenser target RPM and

Conductance Reading, (Diagnostics Only)

Use the (+) or (-) buttons to

adjust % value

NONE

NONE

FlvrGrdRange 10%

(-) EXIT (+)

LfFlvrGrdTargets

RPM XXX-CON XXX

RtFlvrGrdTargets

RPM XXX-CON XXX

For Portion Control Dispensers (LCR-2 PC, LCR-2C PC, LCR-2A PC) only