An interlock is generated and testing stops. Power is removed from
the station.

Interlock

The program (test) stops and an interlock is generated. The station
still has power but the Run button cannot be pressed.

Program Stop
Interlock

The program (test) holds and an interlock is generated. The station
still has power but the Run button cannot be pressed. The test can
be resumed after the interlock is reset.

Program Hold
Interlock

C-Stop Interlock

Note:

The C-Stop Interlock is applicable only to MTS FlexTest
controllers.

A C-Stop (Controlled Stop) Interlock results in the following actions:

A Program Stop Interlock occurs. The test stops but the station
still has power.

The Station Manager C-Stop indicator is turned on.

The Station Manager application logs a message.

The control channel will execute the user-defined C-Stop action,
which is enabled and configured at Station Manager Tools menu
> Channel Options window > C-Stop tab. You must have
Configuration access level to configure the C-Stop Interlock
action.

Note:

The actions of the C-Stop interlock present a fundamental
difference from the Ramp To and Hold At Level user-defined
actions, which require the Function Generator to be in the run
state to execute. C-Stop does not have this requirement.

The program (test) stops but the station still has power.

Program Stop

The program (test) holds with the current command. The station still
has power, and the test can be resumed.

Program Hold

These are custom program actions you can create, such as “Turn on
light.” For more information on using the Station Builder application

Custom Actions

to add resources to a station configuration, see the MTS Series 793
Control Software Manual
.

Effects of Controller Actions

The table below describes how each controller action effects the state of the test, the status of power to
the system, and how the command signal and actuator respond when a limit is tripped.