Landmark Test System Site Prep Guide

16

Load Unit Specifications

System Component Specifications

Environmental Requirements—Series 370 Load Frames

All Series 370 load frames are intended for indoor use only. This indoor
environment must conform to the following environmental specifications.

Note

All Series 370 load frames must only be operated under the installation
and ambient conditions (such as, temperature, moisture, and EMC)
specified.

Electrical Requirements—Series 370 Load Frames

Parameter

Specification

Temperature

5ºC–40ºC (41ºF–104ºF)

Humidity

10–85%, non-condensing

Altitude

2000 m (6,561 ft) maximum

Parameter

Specification

Input Voltage

115 or 230 VAC (Single phase)

Input Frequency

50 or 60 Hz

Input Current

7 A at 115 VAC, 0.4 A at 230 VAC

Environment:

Location: Indoor use only

Operating temperature 5–40°C (41–104°F)

Relative humidity 5-85% non-condensing

Maximum altitude 3000 m (9,842 ft)