Hydraulic Power Unit Specifications

Landmark Test System Site Prep Guide

System Component Specifications

19

Model Specific Specifications

Parameter

505.07 Specification

505.11 Specification

Pump/motor

Maximum Continuous
pressure

Maximum Flow
capacity

Motor rating

Line voltage starter configuration

21 MPa (3000 psi)

22.7 L/m (6 gpm) at 50 Hz
26.5 L/m (7 gpm) at 60 Hz

11.2 Kw (15 hp) at 50/60 Hz

Wye-Delta starter configuration

21 MPa (3000 psi)

41.6 L/m (11 gpm) at 50/60 Hz

18.6 Kw (25 hp) at 50/60 Hz

Water flow rating
(input temperature)

10.0°C (50°F)

15.5°C (60°F)

21.1°C (70°F)

26.7°C (80°F)

32.2°C (90°F)

Heat load (maximum)

3.8 L/m (1.0 gpm)

4.9 L/m (1.3 gpm)

6.1 L/m (1.6 gpm)

8.3 L/m (2.2 gpm)

15.9 L/m (4.2 gpm)

12.3 kW (42,000 Btu/hr)

7.2 L/m (1.9 gpm)

9.1 L/m (2.4 gpm)

12.1 L/m (3.2 gpm)

18.9 L/m (5.0 gpm)

49.2 L/m (13.0 gpm)

20.5 kW (70,000 Btu/hr)

Weight with maximum oil
174 L (46 gal))

450 kg (992 lb)

473 kg (1042 lb)