Page 12

Source/Power & Measure Two Wire Transmitters
Ranges and Resolution

0.000 to 24.000 mA or -25.00 to 125.00% of 4-20 mA

Accuracy

EZ Check(s) at 4 & 20mA

1

0.0 to 24.00 mA

≤ ±(0.025% of Span at 4 mA & 20 mA) (± 0.005mA)
≤ ± (0.05% of 24.00mA Span) (± 0.012mA)

Noise

≤ ± ½ Least Significant Digit

Temperature effect

≤ ± 0.005 %/°C of FS

Loop compliance voltage

≥ 24 DCV @ 20.00mA

Loop drive capability

1200 Ω at 20 mA for 15 hours nominal;
950 Ω with Hart Resistor enabled 334Plus

Battery life

≥ 30 hrs at 12 mA nominal; ≥ 25 hrs with backlight on

1

These are internal calibrated cardinal reference points and accuracy is not defined or limited by display

resolution

2-Wire Transmitter Simulation
Accuracy

Same as Source/Power & Measure

Voltage burden

≤ 2V at 20 mA

Overload/Current limit protection

24 mA nominal

Loop voltage limits

2 to 100 VDC (fuse-less protected from reverse polarity

connections)

Battery life

≥ 125 hours nominal; ≥ 100 hrs with backlight on

Voltage Read
Range and Resolution

-99.99 to +99.99 VDC Full Span (FS)

Accuracy

≤ ± 0.05 % of FS

Temperature effect

≤ ± 100 ppm/°C of FS

Input resistance

≥ 2 MΩ

Battery life

≥ 125 hours nominal; ≥ 100 hrs with backlight on

PIECAL 334Plus Specifications (continued)

(Unless otherwise indicated all specifications are rated from a nominal 23 °C,

70 % RH for 1 year from calibration)

Additional Information

This product is calibrated on equipment traceable to NIST and includes a Certificate of Calibration.

Test Data is available for an additional charge.

Practical Instrument Electronics recommends a calibration interval of one year. Contact your local

representative for recalibration and repair services.