Practical instrument electronics, Reading milliamp outputs, Read dc volts – Practical Instrument Electronics 334Plus User Manual

Page 6

Advertising
background image

Practical Instrument Electronics

82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com

Page 6

Signals exceeding ±60.00 VDC are indicated
by i on the display.

Reading Milliamp Outputs

Milliamp Output Signal

Controller

Transmitter

P/I

DCS

4.334

READ

READ V

Choose this function to measure from -99.99 to +99.99V DC.
1) Select

“READ VDC” with slide switch

q

.

2) Connect the red (+) and black (-) leads of the PIECAL

334Plus across the voltage source to be measured.

Any DC voltage from -99.99 to +99.99 volts may be
measured. Loop power supplies, signal voltages at
receivers, batteries and transmitter voltage drops may be
measured. Signals exceeding ±60.00 VDC are indicated by
i

on the display. Signals exceeding ±99.99 VDC will be

indicated by “OVERRANGE” flashing on the display.

Read DC Volts

Voltage to be Measured

Battery

Loop Power Supply

Transmitter Voltage Drop

1 to 5 Volt Loops

READ

23.34

V

READ

i

72.83

V

READ mA, READ % (Percent of 4 to 20 mA)

Choose this function to measure from 0.000 to 52.000 milliamps or -25.00 to 300.00%.

1) Open the current loop at any convenient point

along the signal path.

2) Select “mA” or “% 4 to 20mA” with slide

switch

q

.

3) Select “Read” using slide switch

w

.

4) Connect the red input lead (+) of the

PIECAL 334Plus to the more positive
point of the break and the black input to
the more negative point.

Signals below 0 mA or open circuits are
indicated by 0.000 mA (-25.00%) on the
display. Signals above 52 mA are current
limited by protection circuitry and
“OVERRANGE” is flashed on the display.

Advertising