General specifications, Rtd simulation specifications, Rtd read specifications – Practical Instrument Electronics 512 User Manual

Page 3: Available options, Model 512 operating instructions, Connection diagrams, Specifications

Advertising
background image

Model 512 Operating Instructions

Connection Diagrams

Four Wire Connection to Transmitter

Specifications

General Specifications:

(Unless otherwise indicated all specifications are rated from a nominal 23 °C, 70 % RH for 1 year from calibration)

Temperature Range

-25 to 60 °C (-10 to 140 °F)

Relative Humidity Range

10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing

10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing

Size

4.9 X 3.15 X 1.82 inches (125.5 X 80 X 46.2 mm)

Weight

9.1 oz (258 grams)

Battery

9V Alkaline provides 45 hours of continuous use

Miscellaneous

Low battery indication with nominal 1 hour of operation left

Protection to 60V for up to 30 seconds in duration

High contrast graphic liquid crystal display with 0.357” (9.07 mm) high digits

Resolution

°C or °F / 0.01 Ω

Span 0.00-410.00

Accuracy

±(0.015 % of Ω + 0.05) Ω (see accuracy tables for temperature error)

Temperature Coefficient

±0.01 % of span in Ω/°C ambient

RTD Simulation Specifications:

Allowable Excitation Current

100 µA to 10.2 mA, steady or pulsed/intermittent/smart

for accuracies below 100µA add

±10µV/Excitation Current (units are in Ω)

Pulsed Excitation Current Compatibility

DC to 0.01 second pulse widths

RTD Read Specifications:

Excitation Current

1.0 mA nominal

Available Options:

Carrying Case

Part Number: 020-0201

841 Holt Road #1 Webster, NY 14580

Tel: 585-872-9350 Fax: 585-872-2638

[email protected] http://www.piecal.com/

Practical Instrument Electronics, Inc. Copyright 2003. All rights reserved.

512-9002 2/25/04

3-4