Model 512 operating instructions, Temperature accuracy, Warranty – Practical Instrument Electronics 512 User Manual

Page 4

Advertising
background image

Model 512 Operating Instructions

841 Holt Road #1 Webster, NY 14580

Tel: 585-872-9350 Fax: 585-872-2638

[email protected] http://www.piecal.com/

Practical Instrument Electronics, Inc. Copyright 2003. All rights reserved.

512-9002 2/25/04

4-4

Temperature Accuracy

The following charts give worst-case temperature accuracy based on stated resistance accuracy of ±(0.015 % + 0.05) Ω.

Pt100 TCR=0.00385

-200 to 850 °C

0.1

0.2

0.3

0.4

-200

0

200

400

600

800

Temperature °C

Error °C

Pt100 TCR=0.003902

-200 to 850 °C

0.1

0.2

0.3

0.4

-200

0

200

400

600

800

Temperature °C

Error °C

Pt100 TCR=0.003916

-200 to 850 °C

0.1

0.2

0.3

0.4

-200

0

200

400

600

800

Temperature °C

Error °C

Pt100 TCR=0.003926

-200 to 850 °C

0.1

0.2

0.3

0.4

-200

0

200

400

600

800

Temperature °C

Error °C

Cu10 TCR=0.00427

-200 to 260 °C

1.2

1.25

1.3

1.35

1.4

-200

-100

0

100

200

300

Temperature °C

Error °C

Ni110 TCR=0.00530

-100 to 300 °C

0.08

0.1

0.12

0.14

0.16

-100

0

100

200

300

400

Temperature °C

Error °C

Ni120 TCR=0.00672

-80 to 260 °C

0.07

0.08

0.09

0.1

-100

-50

0

50

100

150

200

250

300

Temperature °C

Error °C

Warranty

Our equipment is guaranteed against defective material and workmanship (excluding batteries) for a period of three years from the
date of shipment. Claims under guarantee can be made by returning the equipment prepaid to our factory. The equipment will be
repaired, replaced or adjusted at our option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our
guarantee. No responsibility is accepted for damage, loss or other expense incurred through sale or use of our equipment. Under no
condition shall Practical Instrument Electronics, Inc. be liable for any special, incidental or consequential damage.

Advertising