Model 525 operating instructions, Source mode, Auto stepping in source mode – Practical Instrument Electronics 525 Rev F User Manual

Page 3: Rtd connection diagrams, (wire lead kit supplied)

Advertising
background image

Model 525 Operating Instructions

841 Holt Road #1 Webster, NY 14580

Tel: 585-872-9350 Fax: 585-872-2638

[email protected] http://www.piecal.com/

Practical Instrument Electronics, Inc. Copyright © 2004. All rights reserved.

525-9002 5/26/05

3-8

Source Mode

Slide the SIMULATE/OFF/READ switch to SIMULATE for direct RTD or T/C output. The Model 525 outputs resistance corresponding to
temperature for the selected RTD or outputs mV corresponding to temperature for the selected T/C type.

Turn the EZ-Dial™ Knob to change temperature, push and turn for faster dialing.

Slide the EZ-Check™ Switch to HI or LO to recall stored settings. While in the HI or LO position, dial a new setting and press the
EZ-Dial™ Knob to store. The DIAL position always holds the last setting dialed there. Using the STORED Values gives you very
accurate values for repeatability. Adjusting the EZ-Dial™ Knob you can check for controller actions, trip points or hysteresis.

Double-click the EZ-Dial™ Knob to return to the configuration menu.

Auto Stepping in Source Mode

Select the step size by pressing the EZ-Dial™ Knob to find the desired step size or %. Select the step time by pressing the EZ-Dial™
Knob to find the desired step time indicated in seconds.

RTD Connection Diagrams, (wire lead kit supplied)

SOURCE ONLY: (See opposite figure)

To connect to a 2 wire transmitter connect one red lead to
post 1 (RTD1) and connect one black lead to post 2 (RTD2).

To connect to a 3 wire transmitter connect one red lead to
post 1 (RTD1) and connect two black leads to post 2 (RTD2).

To connect to a 4 wire transmitter connect two red leads to
post 1 (RTD1) and connect two black leads to post 2 (RTD2).

NOTE:

3 WIRE & 4 WIRE TERMINALS ARE FOR READ MODE
ONLY. See figure page 2.