RTD Temperature Accuracy

The following charts give worst-case temperature accuracy based on stated resistance accuracy of ± (0.015 % + 0.05) Ω.

Pt100 TCR=0.00385

-200 to 850 °C

0.1

0.2

0.3

0.4

-200

0

200

400

600

800

Temperature °C

Error °C

Pt100 TCR=0.003902

-200 to 648.9 °C

0.1

0.2

0.3

0.4

-200

0

200

400

600

800

Temperature °C

E

rro

r °

C

Pt100 TCR=0.003916

-200 to 648.9 °C

0.1

0.2

0.3

0.4

-200

0

200

400

600

800

Temperature °C

E

rro

r °

C

Pt100 TCR=0.003926

-200 to 850 °C

0.1

0.2

0.3

0.4

-200

0

200

400

600

800

Temperature °C

Error °C

Cu10 TCR=0.00427

-200 to 260 °C

1.2

1.25

1.3

1.35

1.4

-200

-100

0

100

200

300

Temperature °C

Erro

r °C

Cu50 TCR=0.00428

-50 to 150 °C

0

0.1

0.2

0.3

0.4

0.5

-50

0

50

100

150

Temperature °C

Er

ro

r °

C

Ni120 TCR=0.00672

-80 to 260 °C

0.07

0.08

0.09

0.1

-100

-50

0

50

100

150

200

250

300

Temperature °C

Error °C

Ni110 TCR=0.00530

-100 to 260 °C

0.1

Error °C

0.08

0.1

0.12

4

0.16

-100

0

100

200

300

400

Temperature °C