Rp-2009 – RKI Instruments RP-2009 User Manual

Page 16

Advertising
background image

13

RP-2009 Pump for the GX-2009

Replacing the Hydrophobic Disk Filter

and Wire Mesh Disk

1. Locate the clear plastic filter holder at one end of the pump

unit.

2. Grasp the ridged sides of the filter holder and turn it 1/4 of a

turn counterclockwise.

3. Pull the filter holder away from the pump unit. Do not lose

the cotton dust filter.

4. The hydrophobic disk filters and wire mesh disk are located

in the pump unit and are retained by a rubber gasket. Pull
out the rubber gasket with needle nose pliers.

Figure 8: Changing the Hydrophobic & Wire Mesh Filters

5. Remove the old hydrophobic filters and/or wire mesh disk

from the gasket. A hydrophobic filter is located on either side
of the wire mesh disk.

6. Install the new hydrophobic filters and/or wire mesh disk

making sure a hydrophobic filter is located on each side of

POWER

RESET

Rubber Gasket

Installation
Stub

LOW FLOW

POWER

LOCK

Cotton Dust Filter

Hydrophobic Disk Filter

OPEN

RP-2009

RELEASE

Wire Mesh Disk

Filter Holder