19 • Conversion Table Tab

Volt
This value determines the sensor voltage for the catalytic
combustible sensor. It can be set to 1.3 or 2.4. When set to 1.3, the
sensor is in methane elimination mode. When set to 2.4, the sensor
is reading at full response. Even if not all of the user-defined
catalytic combustible channels are being defined, the Volt column
must have a valid number entered. A valid number for the Volt
column is either 1.3 or 2.4.

While the pre-defined gases may not be edited, the 5 user-defined
gases may be edited by doing the following:

1. With the Maintenance Data Loader Program running, click on the

Conversion Data tab and then click on the User-Defined Table tab.

NOTE: Make sure you have downloaded the data from the EAGLE 2

to avoid erasing its data when updating the instrument using
the Transmit New Data button.

2. If no user-defined gases have been previously defined for the

connected instrument, the columns in User-Defined Table will be

Figure 13: User-Defined Table