RKI Instruments 71-0191RK User Manual

Page 30

Advertising
background image

27 • PID Sensor Tab

4. Open the csv file using Notepad, Word, or WordPad. The example

below shows a csv file opened in WordPad. The user defined PID
gas is number 17.

5. The values in the csv file are all separated by commas. These

values are in the same order as the columns in the Maintenance
Data Loader Program. The first value is the gas number, the second
is the gas name, etc.

6. Edit the values you wish to change and save the file.

7. Return to the Maintenance Data Loader Program with the High

Range tab still up and press “Import csv file”.

8. Choose the file you just edited and press “Open”.

9. The values you entered in the csv file will appear in the High

Range tab.

10.If the program finds anything wrong with the values that were

entered, the box containing those values will turn red. Make sure
that you have entered valid characters for each field.

NOTE: The Name and Long Name for the Low Range and High

Range tabs must agree. If they do not, the program will keep
the Low Range Name and Long Name and change the High
Range to agree with it.

11.Repeat step 1-step 10 for the Low Range tab.

12.Once all of the data you wish to enter has been entered and is

Figure 19: CSV File

Advertising