Little Dipper

®

Processor for Agilent— USER MANUAL

July 2013

manual 2.0; software 1.4

17

6. To add a pause within the protocol, do one of the following:

• Enter time for rack to remain submerged at end of step
• Enter 9999 for User Control. Rack will stay submerged until the operator

presses the touchscreen to resume the protocol.
This is beneficial for protocols that require more than 5 baths.

7. When all values for Step 1 are entered, press Save and Exit to return to the

Step Input screen.

8. Continue to enter values for additional steps (max. 9) (Fig. 36). Remember to

program “C” for the centrifuge, if ending in a drying step.

9. When finished creating the protocol, press Save & Exit to return to the Select

Protocol screen. Up to 3 steps are displayed at one time on the touch screen.
Each screen of the protocol must be saved separately before saving the entire
protocol.

B. Deleting a Protocol

1. To delete a protocol, press Control Panel from the home screen then Advanced

Settings.

2. Press the Delete Protocol button to open the screen.
3. Choose a protocol to delete using the up and down arrows and then press

Select. A confirmation screen appears saying “Delete Protocol?” Choose YES to
confirm.

4. Press the BACK button on each of the previous screens to return to the home

screen.

Fig. 36. Advanced Settings screen.

Fig. 37. Delete Protocol button.

Fig. 38. Delete Protocol screen.

Fig. 39. Delete Protocol confirmation.