SciGene Little Dipper Processor for Agilent User Manual

Page 10

Advertising
background image

Little Dipper

®

Processor for Agilent— USER MANUAL

July 2013

manual 2.0; software 1.4

9

C. Computer and Power Connections

1. Connect the braided cable to the touchscreen and to the receptacle labeled

TOUCHSCREEN

” on the right side of the instrument, using the circular connectors

(Fig. 6).

2. Connect the power cord to the receptacle labeled “

POWER

” on the right side of the

instrument (Fig. 7) and to a power source that delivers 15 Amps or more for 115V
units or 8 Amps or more for 230V units. Avoid connecting to a circuit with a
centrifuge, freezer or other equipment that draws significant power.

3. Turn on the main power switch on the right side of the instrument. The

touchscreen should illuminate and the motors should activate (Fig. 8).

4. Briefly turn on power to each of the five baths using the power switch located

above each bath position (Fig. 9). Verify that each temperature controller
illuminates and then turn off power to each bath position.

Baths should not be powered on for more than a few seconds without buffer
in the bath inserts. Do not heat dry baths.

Fig. 6. Connect touchscreen.

Fig. 7. Connect power cord.

Fig. 8. Turn on main power.

Fig. 9. Briefly turn on baths.