SciGene Using Mai Tai Hybridization Cassettes User Manual

User guide, Mai tai, Hybridization system

Advertising
background image

Pub.

406 Rev.E 0611

Mai Tai

®

Hybridization System

SciGene ◊ 617 N. Mary Avenue ◊ Sunnyvale, California 94085 USA ◊ (408) 733-7337 [email protected]

continued on next page...

USER GUIDE

Equipment Needed

Mai Tai

®

Hybridization System*, 115v/230v.

(SciGene Cat. #1057-00-1/1057-00-2)

Includes a Model 777 Microarray Oven, a Mai Tai Rotator
and two Mai Tai Cassettes.

*The Mai Tai Rotator and Cassettes are compatible with

SciGene Series 700 Microarray Ovens (Models 777 and 700)
and Robbins Scientific Model 400 Ovens.

Agilent or other microarrays compatible with Agilent gasket
slides on 1x3 inch (25x75mm) substrates.

Agilent Hybridization Chamber gasket slides, 5-pack, that
match the microarray format used:



Gasket slides, 1

x

244K microarrays

(Agilent cat. #G2534-60003)



Gasket slides, 2

x

105K microarrays

(Agilent cat. #G2534-60002)



Gasket slides, 4

x

44K microarrays

(Agilent cat. #G2534-60011)

Mai Tai

®

Cassette Assembly

Using Mai Tai® Hybridization Cassettes

The cassette has a
standard microplate
footprint so samples may
be loaded manually or with
robotic liquid handling
equipment, such as the
ArrayPrep

®

Target

Preparation System
[SciGene Cat. #2000-00-1
(115v); 2000-00-2 (230v)].

Place the gasket slide down

flat, ensuring the slide does

not rest on the edge of the

base plate (Figure 3). Practice

placing ordinary 1 x 3” (25x75mm) glass slides into the
chamber prior to running an experiment.

4. Pipette labeled sample into the center of each chamber of

the gasket slide using the volumes shown in Table 1.
Load samples onto all gasket slides BEFORE placing arrays.

Mai Tai

®

Hybridization System

Using Mai Tai

®

Hybridization Cassettes

1. With Mai Tai

®

Rotator installed,

pre-heat the hybridization

oven to the desired
incubation temperature.

2. Place the base plate

on a flat surface.

3. Place

a

new gasket

slide into each cassette

chamber with the “Agilent”

label face up and aligned with
the “barcode” label (Figure 2).

Sample loss has been noted

with previously used slides.

The Mai Tai

®

Cassette design uses just two parts that simplify

set up and disassembly and reduce the risk of errors (Figure 1).

Gasket

Format

Probe

Volume

1x microarray

490 µl

2x microarray

245 µl

4x microarray

100 µl

Table 1. Sample Volumes

Don’t touch

the gasket!

DISPENSE SAMPLE IN CENTER

OF GASKET COMPARTMENT

To prevent spilling or leaking, dispense hybridization

solution to the center of the gasket well without touching

the gasket with the pipette tip. Some sample will “wick”

over the barrier if it touches the gasket during assembly.