Using Mai Tai® Hybridization Cassettes

SciGene ◊ 617 N. Mary Avenue ◊ Sunnyvale, California 94085 USA ◊ (408) 733-7337 [email protected]

Using Mai Tai

®

Hybridization Cassettes

(continued)

5. Holding array at the

ends with the array

side down and

ensuring that the

barcode side

matches with the one

on the gasket slide,

bring the array down

slowly holding it

horizontal to the

gasket slide (Figure 5).
Placing the array evenly onto the gasket slide is crucial to
prevent spilling or leakage.
Use Agilent arrays or microarrays from other sources that
are compatible with the Agilent gasket slides.

6. When all arrays have been

placed, make sure all

assemblies are laid flat

against the base plate. See

Figure 6 for examples of

proper and improper
placement.

7. Place the cover plate over

the arrays with the barcode

window matching the

barcodes on array. See Figure 7 for correct placement and
orientation.

11. Rotate the Mai Tai

®

Rotator at the speed and time

recommended by the microarray manufacturer.

12. At the completion of the incubation period, remove the

cassette(s) from the rotator and allow to cool for a few
minutes. Loosen thumbscrews and remove the top plate.

13. Remove all array-gasket-slide assemblies from the cassette

and place them in wash buffer specified by the array
supplier.

Place all assemblies into wash buffer before separating
any arrays from gasket slides.

14. With all assemblies submerged in wash buffer, separate

each array from its gasket slide and place in a slide rack
for processing.

15. Wash and dry the microarray(s) on the Little Dipper

®

Processor [SciGene Cat. #1080-40-1 (115v); 1080-40-2 (230v)]
or manually using the buffer(s) specified by the supplier.

Cleaning and Storage of the Mai Tai

®

Cassette

Hybridization solution may occasionally spill and dry on to the

Mai Tai

®

base and cover plates, making arrays or gasket slides

stick to these parts during hybridization. Base plates, cover

plates and thumbscrews need to be cleaned periodically to

prevent such sticking. Follow the protocol below to clean
Mai Tai

®

parts.

1. Disassemble each Mai Tai

®

cassette (Figure 9) into the base

plate, cover plate and thumbscrews.

Captive thumbscrews can be
unscrewed from the cover plate.

2. Using a lint-free wipe soaked in

70% ethanol, wipe all surfaces of

base plate, cover plate, thumbscrews and thumbscrew

hole carefully to remove dried hybridization solution and
salts.

3. Allow the clean parts to dry completely by evaporation.
4. Assemble the parts back together using the thumbscrews,

cover with cling wrap and store in a dry place.

— End Protocol —

SciGene

Automating Array Workflows

All third party marks are the property of their respective owners.

8. Tighten

the

thumbscrews all the

way to a complete
stop. Don’t use tools.

9. Rotate the cassette

to ensure that the

bubbles move freely

inside each array
chamber.

For optimal hybridization, it is best to have a single, large

bubble which moves freely around the chamber when

rotated. After assembly, ensure there are no stationary

bubbles in any array chamber when the assembly is

rotated. Stationary bubbles can be released by gently

tapping each corner of the assembly on a firm surface as it
is rotated.

10. Place the assembly into

the Mai Tai

®

rotator in

the pre-heated oven.

Raise the spring bar

and load the Mai Tai

®

cassettes into the

sleeves of the rotator as
shown in Figure 8.

When loading multiple cassettes, load equal number of

cassettes on either sides of the rotator to balance weight.

FIGURE 9