Epi light guide installation (geldoc-it, And chemidoc-it – UVP BioLite User Manual

Page 7

Advertising
background image

Manual BioLite MultiSpectral Light Source

7

4. The light guides should be on the inside of the darkroom. Make sure the fiber optic cables extend

from the light guide toward the front of the darkroom (toward you). Match the drilled holes in the
light guide holders to the protruding screws in the epi light structure. Use the brass thumb nuts to
secure the light guides to the darkroom.

5. Remove the rubber cap from the input side of the fiber optic trunk and insert the fiber optic

receptacle into the round silver connector on the front of the BioLite.

6. Secure the fiber optic receptacle to the silver connector with the set screw on the BioLite.


Epi Light Guide Installation (GelDoc-It

®2/TS2

and ChemiDoc-It

®2/TS2

)

1. Remove one stopper at the left side of the darkroom.

2. Remove any existing plates or screws from the mounting bracket so that the epi light guide bracket

appears as follows:

3. Thread the knurled (non-smooth side) of the fiber optic cable from the inside of the darkroom

through the hole created by removing the stopper until the large bulb separating the smooth and
rough side of the fiber optic cable can no longer be pushed through the opening.

Attach light guides to these
screws at top of darkroom
using the brass thumb nuts