Return procedure, Replacement parts and accessories, Troubleshooting – UVP BioLite User Manual

Page 12

Advertising
background image

Manual BioLite MultiSpectral Light Source

12

5. Insert the replacement EKE halogen 150 watt bulb, ensuring that it is fully seated.

6. Replace the cover and secure with the brass nut before plugging unit back in or turning it on.

Return Procedure

A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service before
returning any product. See “Technical Support” below for contact information.

Replacement Parts and Accessories

To order accessories or replacement parts for the BioLite MultiSpectral Light Source, contact UVP’s offices.

Part Description

Part Number

Type EKE 150 Watt Halogen Bulb

34-0088-01


370 – 430 nm Filter (for BFP, CFP, Etc.)

38-0319-01

422 – 478 nm Filter (for GFP, Cy2, Etc.)

38-0319-02

500 - 680 nm Filter (for Cy 5.5)

38-0319-07


Blue Filter (to enhance red and yellow objects)

38-0319-03

Red Filter (darkens green and blue objects)

38-0319-04

Green Filter (increases for red or blue objects)

38-0319-05

Yellow Filter (safelight for photo resist materials)

38-0319-06


Other standard and custom filters are available for a variety of imaging applications. Contact UVP

or visit www.uvp.com for further information.

Troubleshooting

Bulb Does Not Illuminate

1. Recheck the main power cord connection to the BioLite unit.

2. Check the fuses located at the back of the unit next to the power port. A small flat-head screwdriver

or similar tool will be required. Turn the fuseholder cap counterclockwise and the fuse holder will
pop out. Inspect the thin wire within the glass fuse to see if there is a break in the wire. If so,
replace the fuse(s). If fuses are blowing repeatedly, contact UVP Technical Support for additional
troubleshooting.

3. The bulb may be out of position. Reseat the bulb to reestablish the electrical connection by

following the “Bulb Replacement” instruction earlier in this manual.

4. The bulb may be burned out. Follow the “Bulb Replacement” instruction earlier in this manual.

.