UVP HB-500 Minidizer User Manual

Page 15

Advertising
background image

HB-500 Minidizer Hybridization Oven

15

Protocol 3:

Chemiluminescence Detection: HRP-Tagged, Alkaline Phosphatase (AP) Probes
or Antibody Conjugates

Equipment

Clear plastic cling-wrap or Clear transparent sheet protector

UVP EC3 or AC1 Darkroom with Cooled CCD camera

Pipette

Reagents for Chemiluminescence

ECL™ (or other) detection reagent 1

ECL™ (or other if required) detection reagent 2

Membrane following hybridization.

Procedure

1. Mix equal volumes of detection reagents 1 and 2

2. Pipette the mixture over the surface of the membrane and leave at room temperature for 1 minute.

3. Drain the sample and wrap it transfer to Darkroom

4. Close the darkroom and image using CCD camera.



References

Ross, J; Nucleic Acid Hybridization: Essential Techniques; 1998, John Wiley and Sons, ISBN 0-471-97125-1

Sambrook, J; Molecular cloning: a Laboratory manual; 1987, Cold Spring Harbor Laboratory Press; ISBN 0-87969-
309-6

Current Protocols in Molecular Biology; 1987, John Wiley and Sons; ISBN 0-471-50338-X