Service Procedures

Return Procedure

A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service before
returning any product.

Troubleshooting

No Power to the Darkroom

1. Recheck main power cord connections to the darkroom.

2. Check fuses, located at the back of the unit, near the power port. A flat-head screwdriver will be

required. Turn the fuseholder cap counterclockwise and the fuse holder will pop out. Inspect the
thin wire within the glass fuse to see if there is a break in the wire. If so, replace the fuse(s). If fuses
are blowing repeatedly, contact UVP Technical Support for additional troubleshooting.

Transilluminator Will Not Turn On

1. There is a UV interlock switch that turns off the transilluminator and epi UV illumination when the

darkroom door is opened. Be sure the darkroom door is completely closed.

2. Be sure the darkroom cabinet’s “Power On” indicator light is lit. If not, refer to “No Power to the

Darkroom” above.

3. The transilluminator itself has a power switch; make sure that the transilluminator power switch is in

the “On” position.

4. Ensure that the transilluminator power jumper cord is securely connected at both the ends.

5. There is a timer that turns off the transilluminator automatically after 20 minutes of use. Use

VisionWorksLS to deactivate and restart the transilluminator.

Replacement Parts and Accessories

Replacement parts and accessories part numbers are shown below. To order accessories or replacement
parts, contact UVP’s offices listed under Technical Support.

Part Description

Part Number

Power Cord, 100V/115V

46-0023-38

Power Cord, 230V

46-0023-39

Tube, 13W PL White (Sold Individually)

34-0093-01

Tube, 13W BLB 365nm Longwave (Sold Individually)

34-0093-02 (Qty. 2 Required)

Visi-Blue Converter Plate, 21x26cm

38-0200-01

Visi-Blue Converter Plate, 25x26cm

38-0200-04

Thermal printer, digital archive quality 256-grayscale (Mitsubishi)

89-0069-06 (115V)

Thermal printer, digital archive quality 256-grayscale (Mitsubishi)

89-0069-07 (230V)

Thermal printer, digital archive quality 256-grayscale (Sony for Europe) 89-0069-15 (230V)
Thermal paper, Mitsubishi (4 rolls – 800 images)

89-0038-01

Thermal paper, Sony high gloss (5 rolls – 1000 images)

89-0174-01

Thermal paper, Sony glossy (5 rolls – 1000 images)

89-0031-01


Gel-Cutter

85-0002-01

Gel-Ruler

85-0003-01

Gel-Scooper

85-0006-01

Gel-Tray, small

85-0007-01

Gel-Sentry DNA Preparation Plate

97-0076-01

Fluorescent Standard Step Tablet

33-0014-02

Spectacles, UV Blocking (UVC-303)

98-0002-01

Goggles, UV Blocking (UVC-503)

98-0002-02

Faceshield, UV Blocking (UVC-803)

98-0002-04