Introduction

The high sensitivity cooled CCD MegaCam 800

Camera is designed for use with UVP’s

BioSpectrum Imaging Systems. The instruction manual covers assembly of the camera and lens to
the BioSpectrum darkroom and installation of software.

Refer to the BioSpectrum Manual for additional system information and the VisionWorksLS Software
PDF Manual on the Support CD or Help Files embedded in the software for more information.

Applications the MegaCam 800Camera is designed for:

Luminescence detection

Chemiluminescent samples including westerns, northerns, southerns and in vivo/in vitro samples

Autoradiograph films

Fluorescent and colorimetric gels, colony and microtitre plates, dot blots, membrane arrays

Components


Camera Kit


The camera kit includes a motorized lens. Refer to the packing slip or installation components list for a
complete list of parts.

MegaCam 800 Camera

Cables

Power supply for the camera

Motorized lens

Brackets/screws for each lens

Back Focus Rings (Note: Back focus rings may not be required for your camera. If no rings are
needed there will not be any rings shipped with your order.)

Additional components as described in the packing slip/components list

The MegaCam 800 Camera specifications include:


Camera Specifications

MegaCam 800

Type

80mm x 13.5mm full frame CCD

Resolution

3326 x 2504

Pixel Size

5.4µm x 5.4µm

Full Well Capacity

25,000 electrons

Bit Depth

16-bit

Readout Noise

10 electrons (typical)

Quantum Efficiency

(62% @ peak) 400nm 40%;

500nm 58%; 600nm 58%

Cooling

-5

0˚ C from ambient

Binning

Yes

PC Interface

USB

Lens Mount

T-Mount

Lens

50mm f /1.2 or

30mm f/1.4 or

28mm f/1.8 or

Lens Controller

Control box

AC power

USB cable

USB driver