6. Insert the fiber optic bundle into the silver

connector.

7. Attach the cable to the silver connector with the

screw.

Automated BioLite (Software Controlled)

Filters will be installed in their black casing at the UVP factory.

1. To insert the filter/holder into the filter wheel, hold

the assembly so that it is positioned vertically with
the ridge to the top left as shown in the photo.

2. The filter/holder ridge is then positioned closest to

the front of the unit.


3. Slide the holder into the wheel slot.

4. Load the GFP filter in position number one.

5. Load the RFP filter in position number two.

NOTE: Do not force the filter assembly inside the BioLite
wheel slot. First check to ensure that the orientation of
the filter assembly is correct.

Ridge

Side of
Unit