iBox Scientia Small Animal Imaging System

15

Service Procedures

Return Procedure

A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service before
returning any product.

Replacement Parts and Accessories

Replacement parts and accessories part numbers are shown below. To order accessories or replacement
parts, contact UVP’s offices listed under Technical Support.

Part Description

Part Number

Bulb, Type EKE 150 Watt Halogen (for BioLite)

34-0088-01

Power Cord, 100V/115V

46-0023-38

Power Cord, 230V

46-0023-39


Thermal printer, digital archive quality 256-grayscale (Mitsubishi)

89-0069-06 (115V)

Thermal printer, digital archive quality 256-grayscale (Mitsubishi)

89-0069-07 (230V)

Thermal printer, digital archive quality 256-grayscale (Sony for Europe) 89-0069-15 (230V)
Thermal paper, Mitsubishi (4 rolls – 800 images)

89-0038-01

Thermal paper, Sony high gloss (5 rolls – 1000 images)

89-0174-01

Thermal paper, Sony glossy (5 rolls – 1000 images)

89-0031-01

Replacement Instructions

Bulb Replacement for the BioLite

The Type EKE 150 watt halogen bulb has a standard life of 200 hours. It must be replaced with the
same type bulb only. The UVP part number is 34-0088-01. The replacement procedure is as follows:

1.

After allowing the unit sufficient time to cool after last
use (at least a half of an hour), unplug the BioLite
from the wall. Then, use a wrench to open the bulb
access port on the bottom of the BioLite.

2.

Once the nut and access port cover have been
removed, take out the bulb using the mechanism built
into the bulb holder.

3.

Grasp the chrome wire to the left of the bulb, and
push it in the direction of the base of the bulb.

4.

Return the bulb removal mechanism to its original
downward, position.

5.

Insert the replacement EKE halogen 150 watt bulb,
making sure it fully seats.

6.

Replace cover and secure with nut before plugging
unit back in or turning it on.