Care and use manual – Waters XSelect CSH HPLC Columns User Manual

Page 8

Advertising
background image

[ CARE AND USE MANUAL ]

8

XSelect CSH HPLC Columns

5. Measure the peak width at 4.4% of peak height (5-sigma method):

5-sigma Band-spreading (µL) = Peak Width (min) x Flow Rate (mL/min) x

(1000 µL/1 mL)

System Variance (µL

2

) = (5-sigma band-spreading)

2

/ 25

Figure 7: Determination of System Band-spreading Volume Using 5-Sigma Method

In a typical HPLC system, the Band-spreading Volume should be
no greater than 100 µL ± 30 µL (or Variance of 400 µL

2

± 36 µL

2

).

In a microbore (2.1 mm i.d.) system, the Band-spreading Volume
should be no greater than 20 to 40 µL (or Variance no greater
than 16 µL

2

to 64 µL

2

).

c. Measuring Gradient Delay Volume (or Dwell Volume)

For successful gradient-method transfers the gradient delay
volumes should be measured using the same method on both HPLC
systems. The procedure below describes a method for determining
the gradient delay volumes.

1. Replace the column with a zero dead volume union.

2. Prepare mobile phase A (pure solvent, such as methanol) and

mobile phase B (mobile phase A with a UV absorbing sample,
such as (v/v) 0.1% acetone in methanol).

3. Equilibrate the system with mobile phase A until a stable

baseline is achieved.

4. Set the detector wavelength to the absorbance maximum of the

probe (265 nm for acetone).

5. Program a 0-100% B linear gradient in 10 min at 2 mL/

min (the exact conditions are not critical; just make sure the
gradient volume is at least 20 mL) with a hold at 100% B.

Figure 8: Determination of Gradient Delay Volume

6. Determine the dwell time by first locating the time at the midpoint of

the formed gradient (t

1/2

) (half the vertical distance between the initial

and final isocratic segments as shown in Figure 8).

7. Subtract half the gradient time (1/2

tg

) (10 min/2 = 5 min in this

example) from the gradient midpoint (t

1/2

) to obtain the dwell time (t

D

).

8. Convert the dwell time (t

D

) to the dwell volume (V

D

) by multiplying by

the flow rate (F).

Dwell Volume V

D

= (t

1/2

- 1/2

tg

) x F

For fast gradient methods, the gradient delay volume (or dwell
volume) should be less than 1 mL. If the gradient delay volume
is greater than 1 mL, see System Modification Recommendations
section on how to reduce system volume.

VII. ADDITIONAL INFORMATION

a. Use of Narrow-Bore (<3.0 mm i.d.) Columns

This section describes how to minimize extra column effects
and provides guidelines on maximizing the performance of a
narrow-bore column in an HPLC system. A 3.0 mm i.d. narrow-
bore column usually requires no system modifications. A 2.1 mm
i.d. column, however, requires modifications to the HPLC system
in order to eliminate excessive system band-spreading volume.
Without proper system modifications, excessive system band-
spreading volume causes peak broadening and has a large impact
on peak width as peak volume decreases.

System Volume

4.4 %h

5

Advertising