*Chromatographic Conditions:
Column:

2.1 mm x 100 mm,

Sample:

Tryptic digest of cytochrome c;

Flow rate:

0.20 mL/min

Mobile phase A: 0.045% TFA in water;
Mobile phase B: 0.045% TFA in acetonitrile;
Gradient:

from 0 to 15% B in 6 min, from 15 to

36% B in 20 min;

Temperature: 35 ˚C; Detection: UV 214 nm

VI. REP RESENTATIV E T EST CHROMATOGRAPH AND CONDITIONS FOR SEPARATION/

OF P ROT EIN DIGEST

Peak Identification

T1

N-AcGDVEK

T8

TGPNLHGLFGR

T13-T14

KYIPGTK

T15

MIFAGIK

T14

YIPGTK

T5

CAQCHTVEK (heme attached)

T4

IFVQK

T19

EDLIAYLK

T9-T10

KTGQAPGFSYTDANK T12-T13

GITWGEETLMEYLENPKK

T10

TGQAPGFSYTDANK

T12

GITWGEETLMEYLENPK

T19C

EDLIAY