4. Operating procedures

Operating procedures (quick auto stop operation)


Used when you want to “stop fixed temperature operation being performed automatically in several
hours. Quick auto stop operation is a function to enable auto stop timer setting during operation.

Procedures for quick auto
stop operation

MEASURED TEMP.

HEATER

RUN

SET TEMP.

RUN
STOP

TIMER

SUB MENU

1. Setting time period before stop during fixed

temperature operation

Make sure that the RUN lamp is illuminated to indicate

the unit is in operation.
Press the TIMER key.
Characters AStP

are indicated on the

measured temperature screen to indicate the auto
stop operation mode and set duration blinks on the set
temperature screen.

Set a duration you want using the ▼▲ keys.

About the timer function

The maximum time that can be set for the timer is 999 hours
50 minutes.
Up to 99 hours 59 minutes, time can be set in minutes.
One hundred hours and over are set only in 10 minutes.
Keep the ▼▲ keys pressed to continuously change set time
and you can quickly reach the time you want. Press the ▼▲
keys once at a time for fine adjustment.

MEASURED TEMP.

HEATER

RUN

SET TEMP.

RUN
STOP

TIMER

SUB MENU

2. Starting timer operation
When the time you want is set, press the RUN/STOP key

while the set temperature screen is blinking.

The RUN lamp blinks and timer operation is started.
Timer starts counting when the temperature in the bath

reaches the set temperature.

Once timer counting is started, the set temperature screen

changes to the remaining time display.

MEASURED TEMP.

HEATER

RUN

SET TEMP.

RUN
STOP

TIMER

SUB MENU

3.Stopping and ending timer operation
Operation stops automatically when the set temperature

has elapsed.

Characters End

blink on the set temperature screen

to indicate operation has ended.
Press the RUN/STOP key for approx. one second to end
the timer operation mode. The screen switches to the initial
setting screen.