Operating procedures, Using the power outage compensation function – Yamato Scientific DX602 Constant Temperature Drying Oven User Manual

Page 27

Advertising
background image

25

4. Operating procedures

Useful function (power outage compensation function)

Using the power outage
compensation function

The power outage compensation function returns the main
unit operation to the resume status after recovery from power
outage, or keeps the current stop status.

The function is set at “on” on shipping from the factory.

MEASURED TEMP.

HEATER

RUN

SET TEMP.

RUN
STOP

TIMER

SUB MENU

Press the TIMER key (SUB MENU key) long to enter

the sub menu mode.
Press the TIMER key (SUB MENU key) several times
to select the characters Pon

that indicate the

power outage compensation function.

MEASURED TEMP.

HEATER

RUN

SET TEMP.

RUN
STOP

TIMER

SUB MENU

“On” is displayed on the set temperature screen. The

device keeps stop status after recovery from power
outage when this setting is set to “off” using the ▼
key.

Press the TIMER key (SUB MENU key) long to exit
the sub menu mode.

Advertising
This manual is related to the following products: