Description and Function of Each Part

Piping System View

① Water supply port

⑨ Heater

② Solenoid valve

⑩ Heater integument pipe

③ Fixed flow rate valve

⑪ Boiler

④ Micro switch (Heater control)

⑫ Condenser

⑤ Float

⑬ Float switch

⑥ Float tube

⑭ Tank

⑦ Drain port

⑮ Water cock

⑧ Manually cock ( Boiler drain)

Mechanism

1) Water Supply (cooling water, boiler water)

Raw water is controlled to a specified flow rate (about 0.8 liter/min.) by the fixed flow rate
valve, and heat-exchanged in the condenser ⑫ as the cooling water, and then supplied to the
boiler ⑪ via the float tube ⑥, thus overflowing if exceeding the specified level finally to be
drained to outside.

2) Distillation

When the boiler ⑪ reaches the specified water level, the float ⑤in the float tube ⑥ goes up,
and detects the water level, and the heater⑨is turned ON to start distillation. The water
boiled in the boiler turns to steam, cooled/condensed in the condenser ⑫, and then pooled in
the tank as distilled water.

3) Full Tank

When the tank ⑭ reaches the specified amount of water of about 20 liter, the float switch ⑬
trips, and the solenoid valve (for supplying water) ② turns OFF, preventing the water supply
to the main unit, while the heater ⑨ turns OFF to stop distillation. When any specified
amount water has been sampled from the tank, distillation will be restarted.

4) Sampling Water

The distilled water pooled in the tank ⑭ can be sampled by opening the water cock ⑮.