Maintenance method, Warning, Caution – Yamato Scientific WS220 Water Purifiers User Manual

Page 19: Daily inspection and maintenance

Advertising
background image

15

Maintenance Method

Daily Inspection and Maintenance

For the safety use of this unit, please perform the daily inspection and maintenance without fail.
Since this unit distills the city water directly, scale is deposited in the boiler extremely. Therefore,

maintenance and checks are mainly planned to remove such boiler scale.

Reference for maintenance

Maintenance/check items

Reference for timing

Remarks

Washing Distiller

3 months

Washing water supply hose

filter

6 months

Wash the distiller more often in

case the quality of raw water is

not good.

Replacement of hoses

2 years

Connections shall be checked

once per month.

Replacement of flow washers

If the amount of cooling water
decreases and if steam comes

out from the condenser or float

tube.

The flow washer (rubber) in the
fixed flow rate valve deteriorates

in about 2 years. In such a case,

ask for services.

WARNING!

• Disconnect the power cable from the power source when doing an inspection or maintenance

unless needed.

• Perform the daily inspection and maintenance after returning the temperature of this unit to the

normal one.

• Do not disassemble this unit.

CAUTION!

Main unit maintenance

• Use a well-drained soft cloth to wipe dirt on this

unit. Do not use benzene, thinner or cleanser for
wiping. Do not scrub this unit. Deformation,
deterioration or color change may result in.

Monthly maintenance

• Check the earth leakage breaker function.

1. Connect the power cord.
2. Turn the breaker on.
3. Push the red test switch by a ballpoint pen etc.
4. If there is no problem, the earth leakage breaker will be turned off.




For any questions, contact the dealer who you purchased this unit from, or the nearest sales