3 appendix, 1 technical data 3.2 warranty and certificates, 1 warranty 3.2.2 maintenance 3.2.3 certificate – Metrohm 773 pH/mV Simulator User Manual

Page 13: 3 scope of supply and ordering information, 1 technical data

Advertising
background image

3.1 Technical data

773 pH/mV Simulator

11

3 Appendix

3.1 Technical data

Simulated pH values

Range........................................................................................................ pH = 0...20

Divisions......................................................................................................... ∆pH = 1

Temperatures for pH compensation .............................. 0, 20, 25, 40, 60, 80, 100 ° C

Simulated potential values
Range

Divisions

U = 0...20 mV

∆U = 1 mV

U = 0...200 mV

∆U = 10 mV

U = 0...2000 mV

∆U = 100 mV

Turn-on transient

1...10 minutes after switch-on: typically 25 x 10

-6

Influence of ambient temperature
Range

Typically

U

=

±0...20 mV

100 x 10

-6

/K

U

=

±0...200 mV

20 x 10

-6

/K

U

=

±0...2000 mV

15 x 10

-6

/K

Longterm stability (2 years)

Voltage source (pH/mV) .................................................................................. < 0.3 ‰

Resistance for temperature check ..................................................................... < 2 ‰

Source resistances

at U = 0 mV or pH = 7...................................................................................... < 1 Ω

Ri = 1000 MΩ ± 20 %

Ri

∼ 0.002 MΩ between

-20 ... +20 mV............................................max. 2.2 kΩ

-200 ... +200 mV..........................................max. 1.8 kΩ

-2000 ... +2000 mV.........................................max. 600 Ω

Power supply

4 Mignon 1.5V-LR6 batteries; 2200 mAh

Working life 1-2 years

Automatic switch-off after 50 minutes

In order to avoid incorrect measurements resulting from exhausted batteries the
pH/mV Simulator switches itself off when the voltage is insufficient.

Ambient temperature

Nominal working range ............................................................................. 0...+50 ° C

Storage and transport............................................................................. -40...+70 ° C

General construction

Safety specifications according to IEC 1010/EN61010-1, class 3

IP

40

Dimensions
Width

225

mm

Height

90

mm

Depth

95

mm

Weight .......................................................................................................................1.5 kg

Advertising