2.1. Aligning/Checking the potential U/mV

773 pH/mV Simulator

4

2.1 Aligning/Checking the potential U/mV

773 pH/mV Simulator

x100

x10

x1

-

-

14,3k

PT1000

PT100

RE

WE

WE

ON/OFF

pH

U/mV

-

+

20

6

4

2

18

16

14

12

10

8

0

+

+

0

20

25 40 60

80

100

C

1000

~0,002

pH=7

U=0mV

i

R /M

1. Connect the socket "WE" of the 773 pH/mV Simulator with the

measuring input of the instrument with the 6.2150.040 Cable.

For instruments with an E-plug (e.g. 670, 636) use the 6.2150.030 Cable.
For checking the instrument and its electrode cable:

Screw off the cable from the sensor and insert it in socket "WE" of the 773

pH/mV Simulator.

If you normally work with separate electrodes then remove the cable from the

reference electrode and insert it in socket "RE" of the 773.

2. Switch on the 773 pH/mV Simulator; the pilot lamp lights up.
3. Set the instrument to be checked to "Measure U".
4. Select the required factor and sign for U (switch 6).

Example with measurement switch on "2":

Switch setting

Measured value

+ x1

2 mV

- x100

-200 mV

5. Set impedance switch R

i

to 0.002 M

.

6. Set the required value with the measurement switch and read off

the measured value on the instrument being checked.
Compare this measured value with the value given in the table.
Please take into account both the resolution of your instrument’s
display and the information about tolerances given in the
‘Instructions for Use’ for your instrument.

7. Set impedance switch R

i

to 1000 M

. The measured value on the

instrument being checked should differ from the value obtained in
step 6 by less than 1 mV.

Examples:

Set on pH Simulator

Read off on instrument

1500 mV, Ri=

0.002 M

1495 mV

1500 mV, Ri= 1000 M

1494.9 mV

alignment poor, high-impedance OK.

Set on pH Simulator

Read off on instrument

1500 mV, Ri=

0.002 M

1500 mV

1500 mV, Ri= 1000 M

1494.9 mV

alignment OK, high-impedance poor. Check the electrode cable!

Note: the value 0 mV can also be set directly with switch (9) (setting 0 mV).

6.2150.040 Cable