12 pulsation damper, Figure 20, Connecting inline filter – Metrohm 850 Professional IC Anion MCS User Manual

Page 47: See chapter 2.12

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

2 Installation

850 Professional IC – Anion – MCS

■■■■■■■■

37

Installing the inline filter

Caution

Observe the flow direction marked on the filter housing for the connec-
tion of the inline filter.

2

3

2

4

1

Figure 20

Connecting inline filter

1

Connection capillary
Connects the purge valve with the inline fil-
ter

2

PEEK pressure screws, short
6.2744.070

3

Inline filter 6.2821.120
Protects against particles.

4

Connection capillary
Connects the inline filter with the pulsation
damper.

1 Screw on the connection capillary running from the purge valve with

a 6.2744.070 pressure screw to the input side of the inline filters.

2 Screw on the connection capillary running to the pulsation damper

with a 6.2744.070 pressure screw to the output side of the inline fil-
ter.

2.12

Pulsation damper

Note

The pulsation damper is already installed in the newly delivered instru-
ment.

Caution

The pulsation damper is maintenance-free and may not be opened.

Advertising