Metrohm 930 Compact IC Flex Oven/ChS/PP/Deg User Manual

Page 104

Advertising
background image

6.1

■■■■■■■■■■■■■■■■■■■■■■

96

■■■■■■■■

930 Compact IC Flex Oven/ChS/PP/Deg (2.930.2360)

Problem

Cause

Remedy

Guard column – Blocked.

Replace the guard column (see Chapter 3.19,
page 49)
.

Separation column –
Blocked.

Regenerate the separation column (see
Chapter 5.16.4, page 93)
.

Replace the separation column (see "Con-
necting the separation column", page 53)
.

Note: Samples should always be microfiltered .

Injection valve – Valve
blocked.

Have the valve cleaned (by a Metrohm service
technician).

The retention times
in the chromato-
gram have changed
unexpectedly.

Separation column –
Diminished separating effi-
ciency.

Regenerate the separation column (see
Chapter 5.16.4, page 93)
.

Replace the separation column (see "Con-
necting the separation column", page 53)
.

The eluent contains gas
bubbles.

Ensure that the connectors for the eluent
degasser are firmly connected .

Deaerate the high-pressure pump (see Fig-
ure , page 48)
.

High-pressure pump –
Defective.

Request Metrohm Service.

Peak areas are
lower than
expected.

Sample – There is a leak in
the sample path.

Find and fix the leak in the sample path.

Sample – The sample path
is blocked.

Find and fix the blockage in the sample path.

Sample – The sample loop
is not (completely) filled.

Adjust the transfer time for the sample (see
"Determining the transfer time", page 91)
.

Sample – There are gas
bubbles in the sample.

Use the sample degasser .

The peristaltic pump
is pumping too little.

Peristaltic pump – Contact
pressure too weak.

Correctly set the contact pressure .

Peristaltic pump – Filter
blocked.

Replace the filter (see "Replacing the filter",
page 90)
.

Peristaltic pump – Pump
tubing defective.

Replace the pump tubing .