4.15 Initial start-up
■■■■■■■■■■■■■■■■■■■■■■
30
■■■■■■■■
930 Compact IC Flex SeS/PP/Deg (2.930.1560)
MagIC Net detects the instrument and all of its modules.
3 Starting equilibration
■
Start the equilibration in MagIC Net: Workplace
▶ Run ▶ Equi-
libration
▶ Start HW.
4 Deaerating the high-pressure pump
Use the syringe to aspirate eluent until there are no more air
bubbles in the eluent.
5 Setting the contact pressure of the peristaltic pump
■
6 Rinsing the instrument without columns
■
Rinse the instrument (without columns) with eluent for 5 minutes.