Figure 12, Distance to sample rack – Metrohm Liquid Handling Station User Manual

Page 20

Advertising
background image

3.2 Mounting the Liquid Handling Station

■■■■■■■■■■■■■■■■■■■■■■

12

■■■■■■■■

Placing the Liquid Handling Station on the guide

1 Attach the Liquid Handling Station to the assembly rail (11-4) with

the clamping fastener (11-3).

2 Secure the Liquid Handling Station temporarily in place with the

screw (11-2).

Aligning the Liquid Handling Station with the rack

1 Loosen the screws (11-1) of the base plate until you can shift the

Liquid Handling Station with the support bracket.

2 Shift the Liquid Handling Station with the support bracket until the

main body is positioned at an even distance (see Figure 12, page
12)
of 0.5 to 1 mm from the edge of the sample rack.

0.5 ... 1 mm

Figure 12

Distance to sample rack

3 Retighten the screws (11-1) to the base plate.