3 Installation

874 Oven Sample Processor

■■■■■■■■

23

Note

You can adjust the vertical position of the safety shield at any time
by loosening the screws. Take care to ensure that is not possible to
reach into the working area of the needles while the instrument is
in operation.

3.10

Mounting the KF titration cell

The following description is provided as an example for the mounting of a
coulometric KF titration cell. The necessary accessory parts are, with the
exception of the 6.2013.010 clamping ring, not included in the scope of
delivery of the 874 Oven Sample Processor.

801 Stirrer

6.2013.010

6.2047.020

6.0345.100

6.1403.030

6.0341.100

6.1464.320

1

2

3

4

5

6

7

Proceed as follows:

1 Insert the magnetic stirrer 801 Stirrer from above by means of the

support rod at the right-hand side of the tower. Connect the cable to
the coulometer.

2 Screw the 6.2013.010 clamping ring tightly to the support rod,

with the indent facing upward.

3 Fix the 6.2047.020 titration vessel holder to the support rod.

Actuate the fixing lever of the holder while doing so.