12 attaching the sample rack, Figure 16, Attaching the rack – Metrohm 874 USB Oven Sample Processor User Manual

Page 35

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

874 Oven Sample Processor

■■■■■■■■

27

3.12

Attaching the sample rack

1

2

Figure 16

Attaching the rack

The turntable of the 874 Oven Sample Processor must be in initial position
before the sample rack can be removed or put into place. To accomplish
this, carry out the [Rack Reset] function in the manual operation of the
control software.

Put the rack into place as follows:

1 Carefully center the rack on the turntable. The guide bolts on the

turntable must engage with the openings in the bottom of the rack.
Tip: hold the rack in such a way that the printed Metrohm logo is
legible horizontally.

2 Carry out the [Rack Reset] function in the manual operation of the

control software.

The rack is moved into starting position. The magnet code of the
rack is read by the instrument during this process. The white arrow in
figure 16 indicates the position of the magnet holder. The six-digit
magnet code is used to identify the rack type. The sample positions
and any special positions on the rack are defined along with the rack
type.

Advertising