2 Installation

771 IC Compact Interface

10

2.4

Connection of detectors

2.4.1 General

information

Before an external instrument is connected to the 771 IC Compact

Interface, the 771 IC Compact Interface must always be switched off

using mains switch

22

!

The 771 IC Compact Interface has two connections 55 and 66 for analog

input signals (see

Fig. 2

). Information about the technical data of these

interfaces is given in section

4.1

.

For installation and startup of external detectors proceed as follows:

1 Switch off all instruments

Switch off 771 IC Compact Interface and all external detectors

using the mains switch.

2 Connect instruments

Connect the instruments to the 771 IC Compact Interface

according to the connection diagrams shown in sections

2.4.2

to

2.4.4

by using the cables named in the diagrams.

Other instrument combinations can be set up by using these

examples as a guide.

3 Switch on all instruments

Switch on 771 IC Compact Interface and all external instru-

ments using the mains switch.

4 Create new system in «IC Net»

Start the «IC Net» program.

Create a new system file with the selected instruments

(details see

IC Net Instructions for Use

).

2.4.2

One system with one detector

Example

Analysis of cations with conductivity detection.

Instruments

2.771.0010 IC Compact Interface

2.732.0010 IC Detector

2.733.0010 IC Separation Center with 1 injector