5 working procedure, Working procedure – Metrohm 832 KF Thermoprep User Manual

Page 21

Advertising
background image

4.5 Working procedure

832 KF Thermoprep, Operation

17

4.5 Working procedure

Warning!

As both the sample vials and the oven become very hot the sample
vials must be removed from the oven by using the crimping pliers. The
sample vials should only be placed on a heat-resistant surface.


In principle two different methods can be used.

For samples with a low water content one (or more) blank values

must be determined and subtracted from the water content determined
for the sample. Sample vials contain a considerable amount of atmos-
pheric humidity, e.g. Metrohm standard vials with a volume of 6 mL can
contain 60…150 µg water. While the determination is being carried out

atmospheric moisture can also enter the coulometer cell as a result of
leaks, etc.


For samples with a high water content blank value determinations
are not necessary as the atmospheric humidity contained in the vial is

negligible when compared with the moisture contained in the sample. If
no blank determinations are carried out then the whole procedure is

simplified and takes less time.

Determinations with the 832 KF Thermoprep basically take part in three
phases:

1. Preparing / Conditioning the system
2. Possibly determining the blank value

3. Determination of the sample

The procedure for determinations with the 756/831 Coulometer is de-
scribed below. A similar procedure can be used with a volumetric KF ti-
trator (e.g. 795 KFT Titrino).

Note

When using either a coulometer or a volumetric KF titrator it is abso-
lutely essential that an extraction time of at least 300 s is observed per

determination. More details are given in the Instructions for Use of the
particular instrument.

Advertising