3.1 Preparing the ultrafiltration cell

■■■■■■■■■■■■■■■■■■■■■■

6

■■■■■■■■

IC equipment: Inline-Ultrafiltration

3 Installation

This chapter describes how to assemble the ultrafiltration cell, how to fas-
ten the ultrafiltration cell to the Sample Processor or in the ion chromato-
graph and how to connect the ultrafiltration cell to the peristaltic pump
and the injection valve.

Depending on the device combination, the ultrafiltration cell may be
placed at different locations. Please observe the following recommenda-
tions:

Devices

Peristaltic pump

Ultrafiltration cell

Sample Processor with
peristaltic pump and ion
chromatograph

on the Sample Processor

on the Sample Processor

Only ion chromatograph,
type Prep 1

on the ion chromatograph

on the ion chromatograph

Only ion chromatograph
with peristaltic pump

on the ion chromatograph

in the detector chamber of
the ion chromatograph
with cell holder
(6.2057.120)

3.1

Preparing the ultrafiltration cell

You have to insert a filtration membrane (6.2714.020) before you can
insert the ultrafiltration cell.