Metrohm Inline-Ultrafiltration User Manual

Page 18

Advertising
background image

3.2 Connecting the ultrafiltration cell

■■■■■■■■■■■■■■■■■■■■■■

12

■■■■■■■■

IC equipment: Inline-Ultrafiltration

Pump tubing LFL (orange/yellow), 3 stoppers (6.1826.320)

Pump tubing LFL (yellow/yellow), 3 stoppers (6.1826.390)

PEEK capillary, 0.5 mm ID / 1 m (6.1831.060)

PEEK capillary, 0.5 mm ID / 70 cm (6.1831.160)

Pressure screw PVDF (6.2744.000)

Coupling olive/UNF 10/32 (6.2744.030)

Pressure screw, short (6.2744.070)

Pump tubing connection with locking nut (6.2744.160)

1 Preparing the pump tubing for the sample

Use the pump tubing with yellow stoppers (6.1826.390) for convey-
ing the sample.

Attach the coupling olive/UNF 10/32 (6.2744.030) to the inlet.

Attach the pump tubing connection with locking nut
(6.2744.160) to the outlet (see chapter "Installing the peristaltic
pump" in the manual for the ion chromatograph or in the man-
ual for the Sample Processor)
.

2 Preparing the pump tubing for the filtrate

Use the pump tubing with orange/yellow stoppers (6.1826.320) for
conveying the filtrate.

Attach the coupling olive/UNF 10/32 (6.2744.030) to the inlet.

Attach the pump tubing connection with locking nut
(6.2744.160) to the outlet (see chapter "Installing the peristaltic
pump" in the manual for the ion chromatograph or in the man-
ual for the Sample Processor)
.

3 Connecting the capillaries to the pump tubing for the sample

Tighten the PEEK capillary (6.1831.160) to the inlet of the pump
tubing with yellow stoppers (6.1826.390) using a pressure screw
(6.2744.070).

Tighten the PTFE capillary (6.1803.060) to the outlet of the pump
tubing with yellow stoppers (6.1826.390) using a pressure screw
(6.2744.070).