background image

■■■■■■■■■■■■■■■■■■■■■■

4 Installation

940 Professional IC Vario ONE/SeS (2.940.1400)

■■■■■■■■

31

4 Deaerating the high-pressure pump

Use the syringe to aspirate eluent until there are no more air

bubbles in the eluent.

5 Rinsing the instrument without columns

Rinse the instrument (without columns) with eluent for 5 minutes.

4.19

Connecting and rinsing the guard column

CAUTION

New guard columns are filled with solution and sealed with stoppers or
caps on both sides.

Before inserting the guard column, ensure that this solution can be
mixed with the eluent being used (follow information from the manu-
facturer).