2.2

Mains connection

Follow the instructions below for connecting to the power supply. If

the instrument is operated with the mains voltage set wrongly and/or

wrong mains fuse there is a danger of fire!

2.2.1

Setting the mains voltage

Before switching on the 782 IC Column Thermostat for the first time,

check that the mains voltage set on the instrument (see Fig. 3) matches

the local mains voltage. If this is not the case, you must reset the mains

voltage on the instrument as follows:

1 Disconnect mains cable

Disconnect mains cable from mains connection plug 12

12 of the

782 IC Column Thermostat.

2 Remove fuse holder

Using a screwdriver, loosen fuse holder 77 and take out com-

pletely.

3 Change mains voltage

Completely remove voltage selection insert 88 by hand, rotate it

through 180° and reinsert it. The required mains voltage (115 or

230 V) must now be visible from the front.

4 Check fuses

Carefully take both fuses out of fuse holder 77 and check their

specifications:

100

…120 V 0.5 A (slow-blow)

Metrohm-No. U.600.0013

220

…240 V 0.25 A (slow-blow) Metrohm-No. U.600.0010

5 Insert fuses

Change both fuses if necessary and reinsert in fuse holder 77.

6 Install fuse holder

Push fuse holder 77 back into the opening of the 782 IC Column

Thermostat by hand until it clicks into place properly.