3 ic column heater, 1 installation in 733 ic separation center – Metrohm 782 IC Column Thermostat User Manual

Page 13

Advertising
background image

2 Installation

782 IC Column Thermostat

10

2.3

IC Column Heater

2.3.1

Installation in 733 IC Separation Center

Proceed as follows to insert the 2.782.0100 IC Column Heater into the

733 IC Separation Center:

1 Switch off all instruments

Switch off 782 IC Column Thermostat and 732 IC Detector

using the mains switch.

2 Insert IC Column Heater

Unscrew the four knurled screws 34

34 from the top rear panel

35

35 of the 733 IC Separation Center and remove rear panel

(see Fig. 5 of 732/733 Instructions for Use).

Position the IC Column Heater in the inner compartment and

insert the cable permanently attached in opening 29

29 or 33

33 of

the rear panel 35

35.

Replace rear panel 35

35 and screw to the 733 IC Separation

Center using the four knurled screws 34

34.

3 Connect column to injector

Mount a 6.2744.014 compression fitting on each side of the

6.1836.000 preheating capillary (procedure see section 2.5.3

of 732/733 Instructions for Use).

Connect the long, free end of the 6.1836.000 preheating

capillary to connection "4" of the injection valve 68

68 (see

Fig. 16 or Fig. 17 of 732/733 Instructions for Use).

Connect the short, free end of the 6.1836.000 preheating

capillary to the inlet end of the separating column 76

76 (note

flow direction).

4 Rinse column

Place a beaker beneath the column outlet.

Switch on 709 IC Pump and rinse column with eluent for ca.

10 min.

Switch off 709 IC Pump.

5 Connect column to detector block or suppressor

module

without suppressor module:

Screw outlet end of separating column 76

76 to the inlet capillary

82

82 permanently mounted on the detector block 81

81.

with suppressor module:

Screw outlet end of separating column 76

76 to suppressor inlet

capillary 89

89 using a 6.2744.010 compression fitting.