3.3 Connecting the 803 Ti Stand

■■■■■■■■■■■■■■■■■■■■■■

10

■■■■■■■■

803 Ti Stand

4 Press the holding bracket downwards in order to fix the titration ves-

sel. The levers of the holding bracket must enclose the pins of the
plastic ring on the titration vessel in order to ensure a secure hold.

5 Adjust the height of the KF titration cell by pressing the locking lever.

It should almost touch the surface of the stirrer. The position can
now be fixed by readjusting the clamping ring.

Once the height of the KF titration cell has been adjusted correctly,
the entire cell can be raised and swiveled to the right as required by
pressing the locking lever.

3.3

Connecting the 803 Ti Stand

Connect the titration stand as follows:

1 Exit the control software.

2 Connect the connection cable of the titration stand to one of the

sockets marked with MSB on the rear of the control device.

MSB

Waste

Solvent

Figure 3

Connecting the 803 Ti Stand

3 Connecting an additional device (e.g. Dosimat or Dosino) to the MSB

connector of the titration stand (optional).